Yayın:
A method for the determination of the kinetic parameters relevant to atom formation processes in electrothermal atomic absorption spectrometry

dc.contributor.authorBozdogan, AE
dc.date.accessioned2026-06-27T12:56:11Z
dc.date.issued1999
dc.description.abstractA method for the determination of the order of release and the activation energy for the atom formation process from one single absorbance signal under non-isothermal conditions in electrothermal atomic absorption spectrometry is presented. A comparison is made with other methods using the same experimental data sets for Cu, Ni, and Au. The deviations from linearity of the Arrhenius plots are discussed. (C) 1999 Elsevier Science B.V. All rights reserved.en
dc.description.urihttps://doi.org/10.1016/s0584-8547(99)00017-8
dc.identifier.doi10.1016/s0584-8547(99)00017-8
dc.identifier.endpage569
dc.identifier.issn0584-8547
dc.identifier.issue3-4
dc.identifier.startpage557
dc.identifier.urihttps://hdl.handle.net/20.500.14981/47856
dc.identifier.volume54
dc.identifier.wos000079928100013
dc.language.isoeng
dc.publisherPERGAMON-ELSEVIER SCIENCE LTD
dc.relation.ispartofSPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
dc.subjectorder of release
dc.subjectactivation energy
dc.subjectArrhenius plots
dc.subjectelectrothermal AAS
dc.subjectCONSTANT-TEMPERATURE
dc.subjectATOMIZATION
dc.subjectDISSIPATION
dc.subjectMECHANISM
dc.subjectENERGY
dc.subjectORDER
dc.subjectTIME
dc.subjectSpectroscopy
dc.titleA method for the determination of the kinetic parameters relevant to atom formation processes in electrothermal atomic absorption spectrometry