Yayın:
An Adapted Procedure for Calibration of the Non-digital Peak Pressure Indicators: Motor Compression Measuring Devices

dc.contributor.authorEkici, C.
dc.date.accessioned2026-06-27T14:02:15Z
dc.date.issued2017
dc.description.abstractThis paper is written for proposing a guideline to calibrate non-digital peak pressure gages. Motor compression testers are peak pressure measuring devices. Accuracy tests were applied to the motor compression testers and the accuracy classes of the devices were found in the study. These accuracy classes were used to select calibration procedure according to Euramet's Guidelines on the Calibration of Electromechanical Manometers and Euramet's procedure was adapted to calibration of non-digital peak pressure measuring devices. About seventy devices were tested in this study to find their accuracy classes. Measurement uncertainty budgets and the adapted procedure were given in this study.en
dc.description.urihttps://doi.org/10.1007/s12647-016-0194-0
dc.identifier.doi10.1007/s12647-016-0194-0
dc.identifier.eissn0974-9853
dc.identifier.endpage94
dc.identifier.issn0970-3950
dc.identifier.issue2
dc.identifier.startpage87
dc.identifier.urihttps://hdl.handle.net/20.500.14981/56615
dc.identifier.volume32
dc.identifier.wos000402818100001
dc.language.isoeng
dc.publisherMETROLOGY SOC INDIA
dc.relation.ispartofMAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA
dc.subjectMotometer
dc.subjectPressure calibration
dc.subjectPeak pressure indicators
dc.subjectMotor compression testing devices
dc.subjectSTANDARD
dc.subjectSTABILITY
dc.subjectMPA
dc.subjectInstruments & Instrumentation
dc.subjectPhysics
dc.titleAn Adapted Procedure for Calibration of the Non-digital Peak Pressure Indicators: Motor Compression Measuring Devices