Yayın:
New Urea Biosensor Based on Urease Enzyme Obtained from Helycobacter pylori

dc.contributor.authorDindar, Bahar
dc.contributor.authorKarakus, Emine
dc.contributor.authorAbasiyanik, Fatih
dc.date.accessioned2026-06-27T13:14:28Z
dc.date.issued2011
dc.description.abstractThe urease enzyme of Helicobacter pylori was isolated from biopsy sample obtained from antrum big curvature cell extracts. A new urea biosensor was prepared by immobilizing urease enzyme isolated from Helicobacter pylori on poly(vinylchloride) (PVC) ammonium membrane electrode by using nonactine as an ammonium ionophore. The effect of pH, buffer concentration, and temperature for the biosensor prepared with urease from H. pylori were obtained as 6.0, 5 mM, and 25 A degrees C, respectively. We also investigated urease concentration, stirring rate, and enzyme immobilization procedures in response to urea of the enzyme electrode. The linear working range of the biosensor extends from 1 x 10(-5) to 1 x 10(-2) M and they showed an apparent Nernstian response within this range. Urea enzyme electrodes prepared with urease enzymes obtained from H. pylori and Jack bean based on PVC membrane ammonium-selective electrode showed very good analytical parameters: high sensitivity, dynamic stability over 2 months with less decrease of sensitivity, response time 1-2 min. The analytical characteristics were investigated and were compared those of the urea biosensor prepared with urease enzyme isolated from Jack bean prepared at the same conditions. It was observed that rapid determinations of human serum urea amounts were also made possible with both biosensors.en
dc.description.sponsorshipYildiz Technical University [28-01-02-15]
dc.description.urihttps://doi.org/10.1007/s12010-011-9348-2
dc.identifier.doi10.1007/s12010-011-9348-2
dc.identifier.eissn1559-0291
dc.identifier.endpage1321
dc.identifier.issn0273-2289
dc.identifier.issue5-6
dc.identifier.pubmed21881954
dc.identifier.startpage1308
dc.identifier.urihttps://hdl.handle.net/20.500.14981/50924
dc.identifier.volume165
dc.identifier.wos000297222400020
dc.language.isoeng
dc.publisherHUMANA PRESS INC
dc.relation.ispartofAPPLIED BIOCHEMISTRY AND BIOTECHNOLOGY
dc.subjectHelicobacter pylori
dc.subjectUrease enzyme
dc.subjectIsolation
dc.subjectUrea
dc.subjectPolyvinylchloride
dc.subjectCarboxylated polyvinylchloride
dc.subjectPalmitic acid
dc.subjectNonactin
dc.subjectION-SELECTIVE ELECTRODES
dc.subjectPOTENTIOMETRIC BIOSENSOR
dc.subjectPVC MEMBRANE
dc.subjectHELICOBACTER-PYLORI
dc.subjectIMMOBILIZED ENZYME
dc.subjectBLOOD
dc.subjectSENSOR
dc.subjectSERUM
dc.subjectCREATININE
dc.subjectBiochemistry & Molecular Biology
dc.subjectBiotechnology & Applied Microbiology
dc.titleNew Urea Biosensor Based on Urease Enzyme Obtained from Helycobacter pylori