Yayın:
Gas-Screen Slotted Quartz Tube Atomic Absorption Spectrometry: A Remedy for Reducing Interference Effects of Calcium and Chromium

Yükleniyor...
Küçük Resim

Tarih

Kurum Yazarları

Item type:Araştırmacı/Yazar,
BAKIRDERE, Sezgin

Danışman

item.page.editor

Editör

Bölüm / Program

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

TAYLOR & FRANCIS INC

DOI

10.1080/00032719.2012.747093

Türü

View PlumX Details

Araştırma Projeleri

Akademik Birimler

Dergi Sayısı

Özet

A simple device for the reduction of nonspectral interferences in flame atomic absorption spectrometry is proposed. It has been reported that the use of a gas screen (GS) system together with a slotted quartz tube (SQT) enhances the residence time of analyte atoms in measurement zone even more than the SQT alone. This combination causes enhancement of sensitivity and improves the reproducibility of absorbance measurements. In addition, it protects the optical windows of the atomic absorption spectrometer. The operational mechanism of gas screen is simply applying two argon gas walls at both ends of SQT to provide an environment that is partly protected from air. This action enhances the sensitivity of measurement. In this study, interference effects of excess amounts of calcium and chromium on the measurements of Cd, Co, Cu, Pb, Mn, Ni, Se, and Zn were studied using flame AAS. The presence of both Ca and Cr cause higher absorbance values; it is suggested that this is due to formation of oxide species of Ca and Cr and as a result analyte oxide production is suppressed. Therefore, analyte free atom population and sensitivity are enhanced. This enhancement results in a positive error in measurements. For instance, presence of solely 8.0mgL1 of Cr or Ca for Pb as the analyte enhances the signals by 75% and 56%, respectively. When SQT or GS-SQT is used, this effect is significantly reduced.

Tanım

Dergi veya Seri

ANALYTICAL LETTERS

ISSN

0003-2719

ISBN

Haklar

Alıntı

Koleksiyonlar

Onay

Gözden geçir

Tamamlayıcı Bilgiler

Referans Gösteren

Related Patent

Related Goal

0

Views

0

Downloads