Yayın:
Calibration of Heat Stress Monitor and its Measurement Uncertainty

dc.contributor.authorEkici, Can
dc.date.accessioned2026-06-27T14:02:17Z
dc.date.issued2017
dc.description.abstractWet-bulb globe temperature (WBGT) equation is a heat stress index that gives information for the workers in the industrial areas. WBGT equation is described in ISO Standard 7243 (ISO 7243 in Hot environments-estimation of the heat stress on working man, based on the WBGT index, ISO, Geneva, 1982). WBGT is the result of the combined quantitative effects of the natural wet-bulb temperature, dry-bulb temperature, and air temperature. WBGT is a calculated parameter. WBGT uses input estimates, and heat stress monitor measures these quantities. In this study, the calibration method of a heat stress monitor is described, and the model function for measurement uncertainty is given. Sensitivity coefficients were derived according to GUM. Two-pressure humidity generators were used to generate a controlled environment. Heat stress monitor was calibrated inside of the generator. Two-pressure humidity generator, which is located in Turkish Standard Institution, was used as the reference device. This device is traceable to national standards. Two-pressure humidity generator includes reference temperature Pt-100 sensors. The reference sensor was sheltered with a wet wick for the calibration of natural wet-bulb thermometer. The reference sensor was centred into a black globe that has got 150 mm diameter for the calibration of the black globe thermometer.en
dc.description.urihttps://doi.org/10.1007/s10765-017-2224-8
dc.identifier.doi10.1007/s10765-017-2224-8
dc.identifier.eissn1572-9567
dc.identifier.issn0195-928X
dc.identifier.issue6
dc.identifier.urihttps://hdl.handle.net/20.500.14981/56622
dc.identifier.volume38
dc.identifier.wos000400936600006
dc.language.isoeng
dc.publisherSPRINGER/PLENUM PUBLISHERS
dc.relation.ispartofINTERNATIONAL JOURNAL OF THERMOPHYSICS
dc.subjectBlack globe temperature
dc.subjectCalibration
dc.subjectDry-bulb temperature
dc.subjectMeasurement uncertainty
dc.subjectNaturalwet-bulb temperature
dc.subjectWet-bulb globe temperature (WBGT)
dc.subjectThermodynamics
dc.subjectChemistry
dc.subjectMechanics
dc.subjectPhysics
dc.titleCalibration of Heat Stress Monitor and its Measurement Uncertainty