Acoustically Levitated Sample Introduction for GC-MS: A Novel Contactless Approach to Direct Volatile Analysis

dc.contributor.authorAkkaya, Erhan
dc.contributor.authorKoyuncu, Ikbal
dc.date.accessioned2026-06-27T15:19:33Z
dc.date.issued2025
dc.description.abstractConventional GC-MS sample introduction methods often involve physical contact, increasing contamination risk and analyte degradation, especially in sensitive matrices. This study introduces a novel, contactless interface using standing-wave acoustic levitation to suspend analytes in mid-air prior to GC-MS analysis. A 28 kHz ultrasonic transducer, powered by a high-voltage amplifier, generates stable levitation nodes inside a custom chamber, enabling droplets and solid particles to remain airborne during volatilization. Volatile analytes are transferred via heated gas mixture directly into the GC inlet. Acetone droplets yielded strong, reproducible m/z 58 ion signals, and real samples like cucumber (Cucumis sativus) skin demonstrated broader applicability. Helium-nitrogen gas mixture was optimized for both levitation stability and ionization efficiency. The system minimized surface contact, preserved sample morphology, and reduced cross-contamination risks. This platform establishes acoustic levitation as a viable sample delivery strategy for GC-MS, offering enhanced cleanliness, reduced sample loss, and novel handling of delicate or trace-level materials. It opens new directions for automation and integration in analytical workflows.en
dc.description.sponsorshipResearch Fund of the Yimath
dc.description.sponsorshipldimath
dc.description.sponsorshipz Technical University [FBA-2021-4566]
dc.description.urihttps://doi.org/10.1002/rcm.10122
dc.identifier.doi10.1002/rcm.10122
dc.identifier.eissn1097-0231
dc.identifier.issn0951-4198
dc.identifier.issue22
dc.identifier.pubmed40792661
dc.identifier.urihttps://hdl.handle.net/20.500.14981/69744
dc.identifier.volume39
dc.identifier.wos001547579400001
dc.language.isoeng
dc.publisherWILEY
dc.relation.ispartofRAPID COMMUNICATIONS IN MASS SPECTROMETRY
dc.rightsopenAccess
dc.subjectacoustic levitation
dc.subjectcontactless sampling
dc.subjectGC-MS
dc.subjectultrasonic transducer
dc.subjectvolatile analysis
dc.subjectBiochemistry & Molecular Biology
dc.subjectChemistry
dc.subjectSpectroscopy
dc.titleAcoustically Levitated Sample Introduction for GC-MS: A Novel Contactless Approach to Direct Volatile Analysis