Yayın: Electrochemical study of the corrosion behaviour of Co-Cr-Mo orthopedic implants
Yükleniyor...
Tarih
Yazarlar
Danışman
item.page.editor
Editör
Bölüm / Program
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
CARL HANSER VERLAG
DOI
Türü
Özet
A rapid growth is taking place in the use of implant material for the organs malfunctioning or injured. For this reason, search is done for material with physical and chemical resistance suitable for replacement of organs. In this study, the corrosion potential of Co-Cr-Mo alloy according to ASTM F75 at metallic implant quality was measured in open atmosphere Flank solution at 6 mV/min scanning velocity. The pH value of the blood changes as a result of infections in the human body and so, experiments were done at pH values of 4,7 and 9. Fitting potentials showed although not much but a little increase towards basic medium. Although in Co-Cr-Mo alloy grain boundary corrosion is caused it has a higher corrosion resistance than 316L austenitic stainless steel.
Tanım
Dergi veya Seri
PRAKTISCHE METALLOGRAPHIE-PRACTICAL METALLOGRAPHY
ISSN
0032-678X