Yayın:
Determination of Glucose using Flow Injection Analysis and Borate Selective Electrode

dc.contributor.authorUner-Bahar, Duygu
dc.contributor.authorIsildak, Ibrahim
dc.date.accessioned2026-06-27T14:33:18Z
dc.date.issued2020
dc.description.abstractThis study focuses on the development of a flow injection analysis method that performs nonenzymatic glucose measurement. It is aimed to develop a more economical blood glucose monitoring system by measuring the instant blood glucose level indirectly with borate based ion selective electrode. A non-enzymatic electrode based on carbon nanotube-silver borate (CNT-SB) composite electrode was developed and integrated into a flow injection analysis system as a detector, and glucose measurement was performed. Potentiometric measurement was made with this detection chamber. Heights of potentiometric peaks obtained were used for the determination of the amount of glucose in blood samples. Repeated injections were tested by using a mini-sized separation column as stationary phase and successful results were obtained. In addition, boronic acid solution was used as a carrier solution in the system, and it is predicted to be more accessible and simple method than the present non-enzymatic methods as it has ability at neutral pH conditions. The developed glucose measurement system was simple, economic and fast.en
dc.description.sponsorshipIstanbul Gelisim University Scientific Research Projects Application and Research Center [DUP_111019_DUB]
dc.description.urihttps://doi.org/10.20964/2020.12.15
dc.identifier.doi10.20964/2020.12.15
dc.identifier.endpage12739
dc.identifier.issn1452-3981
dc.identifier.issue12
dc.identifier.startpage12724
dc.identifier.urihttps://hdl.handle.net/20.500.14981/62002
dc.identifier.volume15
dc.identifier.wos000600769100084
dc.language.isoeng
dc.publisherELSEVIER
dc.relation.ispartofINTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE
dc.rightsopenAccess
dc.subjectBorate Sensor
dc.subjectDiabetes Diagnosis
dc.subjectEconomic Blood Glucose Measurement
dc.subjectHealth quality
dc.subjectSustainability
dc.subjectSENSOR
dc.subjectElectrochemistry
dc.titleDetermination of Glucose using Flow Injection Analysis and Borate Selective Electrode
dc.typeArticle
dspace.entity.typePublication
local.import.sourceWOS

Dosyalar

Koleksiyonlar