Yayın: Radiation dose variability in critical thoracic organs during CT imaging: A multi-centre phantom dosimetry study
Yükleniyor...
Tarih
Danışman
item.page.editor
Editör
Bölüm / Program
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
PERGAMON-ELSEVIER SCIENCE LTD
DOI
10.1016/j.radphyschem.2025.113174
Türü
Özet
This multicenter phantom-based study quantified organ-specific radiation doses delivered during routine thoracic CT using five clinical systems from four major manufacturers. A total of 70 TLD-100 dosimeters were calibrated and strategically positioned in an Alderson Rando (R) phantom to measure absorbed doses in the lungs, heart apex, atrium, and bilateral breasts. Marked dose variability was observed across scanners, driven by differences in tube current, CTDIvol, pitch, and the implementation of dose modulation strategies. Lung doses ranged from 4.60 to 22.51 mGy, heart apex doses from 4.15 to 23.15 mGy, atrial doses from 4.73 to 20.77 mGy, and breast doses from 3.17 to 22.01 mGy. The CT-2 system consistently yielded the highest organ doses, while the CT-1 demonstrated the most effective dose reduction. Comparative assessment with literature confirmed strong alignment of our measured values with published experimental and Monte Carlo-based studies. It can be concluded that significant dose reductions to radiosensitive thoracic organs are achievable through advanced modulation technologies and tailored protocol adjustments.
Tanım
Dergi veya Seri
RADIATION PHYSICS AND CHEMISTRY
ISSN
0969-806X