Yayın: Field-Free Line Magnetic Particle Imaging Magnet Design Using Nested Halbach Cylinders
Yükleniyor...
Tarih
Yazarlar
Danışman
item.page.editor
Editör
Bölüm / Program
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI
10.1109/lmag.2022.3159446
Türü
Özet
Magnetic particle imaging (MPI) is a novel imaging technique that is a promising candidate for practical use in the medical field. The field-free line (FFL) selection field method in MPI provides spatial encoding along a line, resulting in a faster acquisition time and enhanced sensitivity with increased signal-to-noise ratio. To obtain FFL, a magnet system was designed using nested Halbach rings with octagonal-shaped permanent magnets. In this specific study, simulation studies were implemented using this magnet system for a real case. For this purpose, gradient values and stabilities of the magnet system were calculated. In this investigation, a gradient field within 60 mm stability was obtained along each axis. The gradient field attained values up to 6.1 T/m, which is a highly important parameter for spatial resolution in MPI systems.