Yayın: Determination of trace amounts of copper in tap water samples with a calix[4]arene modified carbon paste electrode by differential pulse anodic stripping voltammetry
Yükleniyor...
Tarih
Danışman
item.page.editor
Editör
Bölüm / Program
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
WILEY-V C H VERLAG GMBH
DOI
10.1002/elan.200603829
Türü
Özet
A calix[4]arene modified carbon paste electrode was used for trace determination of copper. The study of the preconcentration of copper as well as the other heavy metal ions at the modified electrode, with subsequent measurement by differential pulse anodic stripping voltammetry (DPASV), indicates the efficient open-circuit accumulation of the analytes onto the electrode. Many parameters such as the composition of the paste, pH, preconcentration time and stirring rate influence the response of the measurement. The procedure was optimized for copper determination. For a 10-minute preconcentration time at pH 6.5 - 7.5, the detection limit (LOD) was 1.1 mu g L-1. The optimized method was successfully applied to the determination of copper in tap water sample by means of standard addition procedure. The copper content of the sample was comparable with the result obtained with AAS method.
Tanım
Dergi veya Seri
ELECTROANALYSIS
ISSN
1040-0397
ISBN
Haklar
Anahtar Kelimeler
calix[4]arene , carbon paste electrode , stripping voltammetry , heavy metals , tap water , CHEMICALLY-MODIFIED ELECTRODES , PRINTED ELECTROCHEMICAL SENSORS , SELECTIVE COMPLEXATION , ANALYTICAL-CHEMISTRY , METAL-IONS , CALIXARENES , LEAD , CADMIUM , ELECTROANALYSIS , CALIX<4>ARENES , Chemistry , Electrochemistry